Figure 0个  Table 3
    • VirusNo. detectedSensitivitySpecificityPPVNPV
      M+R+M+R-M-R+M-R-(%)(%)(%)(%)
      RSV4692629433.39888.576.2
      FluA432240011.199.566.792.6
      hMPV21934057099.387.597.8
      PIV3197540773.198.979.298.4
      H1N1-p**101502894010010095.1
      PIV1126341766.799.38098.6
      hBoV108042055.610010098.1
      OC43411142226.799.88097.5
      EnV2123039414.392.96.397
      ADV43043157.110010099.3
      NL633314315099.87599.3
      229E3214326099.87599.5
      FluB2204345010010099.5
      PIV421143466.799.866.799.8
      PIV2030435NA100NA99.3
      HKU1000438NANANANA
      * RhV was not evaluated as it was not detected by the monoplex real-time RT-PCR. PPV, positive predictive value; NPV, negative predictive value. ** The Resplex Ⅱ Panel v2.0 was replaced by a ResPlex Ⅱ Plus Panel PRE to enhance H1N1-P coverage during the study. A total of 314 samples were tested by the ResPlex Ⅱ Plus Panel PRE version.

      Table 1.  The comparison of ResPlex (R) and monoplex real-time TaqMan RT-PCR (M) for 16 respiratory viruses *

    • Monoplex
      copies/mL
      No. of ResPlex-PCR positive/No. of Monoplex PCR positive
      RSV (%)FluA (%)hMPV (%)PIV3 (%)H1N1-p (%)PIV1 (%)hBoV (%)
      ≥1066/9 (66.7)1/1 (100)3/4 (75.0)2/3 (66.7)4/5 (80.0)0/0 (0)9/11 (81.8)
      1-9.9×10534/45 (75.5)2/5 (40.0)10/14 (71.4)8/8 (100.0)5/6 (83.3)5/5 (100.0)1/3 (33.3)
      1-9.9×1045/38 (13.2)1/7 (14.3)6/9 (66.7)4/5 (80.0)1/6 (16.7)6/7 (85.7)0/2 (0)
      1-9.9×1031/33 (3.0)0/12 (0)2/3 (66.7)2/4 (50.0)0/6 (0)1/3 (33.3)0/1 (0)
      1-9.9×1020/13 (0)0/11 (0)0/0 (0)3/6 (50.0)0/2 (0)0/3 (0)0/1 (0)
      Negative6/300 (2.0)2/402 (0.5)3/408 (0.7)5/412 (1.2)0/289 (0)3/420 (0.7)0/420 (0)
      z-6.859-5.119-1.267-1.011-2.538-1.068-1.511
      p0.0000.0000.2050.3120.0110.2850.131

      Table 2.  Relationship of sensitivity of ResPlex PCR and viral loads determined by monoplex PCR

    • VirusResplex Ⅱ Panel
      No. of specimens
      Sensitivity
      (%)
      ResPlex Ⅱ Plus Panel PRE
      No. of specimens
      Sensitivity
      (%)
      χ2
      value
      P
      value
      M+R+M+R-M-R+M-R-M+R+M+R-M-R+M-R-
      RSV142538235.93267321232.30.1610.688
      FluA22211850.023002826.36.8910.053
      hMPV44211450.0175129177.32.0780.195
      PIV3602116100.0137329165.02.8740.146
      PIV1302119100.096129860.01.800.515
      hBoV51011883.357030241.72.8130.152
      * R, ResPlex Ⅱ Panel; M, monoplex real-time TaqMan RT-PCR.

      Table 3.  Sensitivity of two kinds of ResPlex Ⅱ kit for common viruses *